Options
PMT-scintillator system set up for D-D neutron TOF measurements in INTI plasma focus device
Citation
Damideh, V., Saw, S. H., Sadighzadeh, A., Jalil Ali, Rawat, R. S., Lee, P., & Lee, S. (2017). PMT-scintillator system set up for D-D neutron TOF measurements in INTI plasma focus device. AIP Conference Proceedings, 1824(1), Article 030028. https://doi.org/10.14221/ajte.2017v42n2.3
Author
Damideh, Vahid
•
Saw, Sor Heoh
•
Sadighzadeh, Asghar
•
Jalil Ali
•
•
•
Lee, Sing
Abstract
This paper summarizes a Photomultiplier-Scintillator diagnostic system for use in our plasma focus experiments at the Center for Plasma Research INTI IU. The system features an anode-grounded high pulse linearity voltage divider and uses NE102A plastic scintillators. It has detected D-D neutrons in INTI plasma focus device with clear and high signal to noise ratio. Neutron TOF of 120 ns has been measured from the time difference between hard x-ray pulse peak and neutron peak time over a flight path of 2.6±0.01 m; giving energy of 2.5±0.1 MeV for these side-on neutrons.
Date Issued
2017
DOI
10.14221/ajte.2017v42n2.3