Options
Absolute measurements of fast neutrons using yttrium
Citation
Roshan, M. V., Springham, S. V., Rawat, R. S., Lee, P., & Krishnan, M. (2010). Absolute measurements of fast neutrons using yttrium. Review of Scientific Instruments, 81(8), Article 083506. https://doi.org/10.1063/1.3478020
Author
Abstract
Yttrium is presented as an absolute neutron detector for pulsed neutron sources. It has high sensitivity for detecting fast neutrons. Yttrium has the property of generating a monoenergetic secondary radiation in the form of a 909 keV gamma-ray caused by inelastic neutron interaction. It was calibrated numerically using MCNPX and does not need periodic recalibration. The total yttrium efficiency for detecting 2.45 MeV neutrons was determined to be ƒn∽4.1 x 10-4 with an uncertainty of about 0.27%. The yttrium detector was employed in the NX2 plasma focus experiments and showed the neutron yield of the order of 108 neutrons per discharge.
Date Issued
2010
Publisher
American Institute of Physics
Journal
Review of Scientific Instruments
DOI
10.1063/1.3478020